Page 1 - OPERATION & MAINTENANCE

ENMETENMET CorporationPO Box 979Ann Arbor, MI 48106-097980003-02704/27/95MCN-204, 10/20/98MCN-285, 11/20/02ISA – MISA – MOPERATION & MAINTENANCEMA

Page 2 - List of Figures and Tables

ENMET Corporation ISA – M85.0 Operation5.0 OperationMount and install the ISA – M electronics unit and sensor as outlined in the previous section (4.0

Page 3

ISA – M ENMET Corporation96.0 Routine Gas Test and Recalibration6.0 Routine Gas Test and RecalibrationPerform this test at regular intervals, you deci

Page 4 - OPERATE – Normal Operation

ENMET Corporation ISA – M10Table 5: Approximate Alarm TimeGas or Vapor Time to Alarm20 ppm CO 5 - 7 minutes50 ppm CO 5 minutes200 ppm CO 2 minutes10%

Page 5 - OTE: N.C.= normally closed

ISA – M ENMET Corporation117.0 Maintenance7.0 MaintenanceThe ISA – M gas monitor is a safety instrument and requires periodic testing and calibration

Page 6

ENMET Corporation ISA – M127.3 Replacement Part NumbersENMET replacement part numbers:Description Part NumberMeter 03411-000Lamp body 62012-010Lens, r

Page 7 - Table 3: Same Density as Air

ISA – M ENMET Corporation139.0 WARRANTY9.0 WARRANTYENMET warrants new instruments to be free from defects in workmanship and material under normal use

Page 8 - 4.3 Sensor Hook-up

Table of Contents1.0 INTRODUCTION...

Page 9 - 4.4 Sensor Heater Voltage

ISA – M ENMET Corporation11.0 Introduction1.0 IntroductionThe ENMET Model ISA-M detector/alarm is an all solid-state electronic gas-detecting instrume

Page 10 - 5.0 Operation5.0 Operation

ENMET Corporation ISA – M22.0 Features if the 2.0 Features if the ISA – M2.1 External Feature of the ISA – MFigure 1 shows the ISA-M control unit whic

Page 11 - 6.1 Test

ISA – M ENMET Corporation32.2 Internal Features of the ISA – MFigure 2 shows the circuit board and terminal blocks housed inside the hinged oiltight c

Page 12 - 6.2 To Recalibrate

ENMET Corporation ISA – M44.0 Installation4.0 InstallationThe ISA – M is completely oiltight. Use the correct oiltight fitting for the cord or condui

Page 13 - 7.2 Sensor Replacement

ISA – M ENMET Corporation54.2 Sensor LocationGases have different densities. Some are heavier than air and concentrate at the bottom of a space. Som

Page 14 - Description Part Number

ENMET Corporation ISA – M64.3 Sensor Hook-upThe MOS sensor is connected to the ISA-M control unit with three conductor wiring, use the correct oiltigh

Page 15 - 9.0 WARRANTY9.0 WARRANTY

ISA – M ENMET Corporation74.4 Sensor Heater VoltageHeating the sensor promotes the oxidation reaction on the element surface. The temperature to whic