5.0 Features5.0 Features
Symptoms Cause Solutions
Screen is blank Used primary cell or discharged battery Charge the primary cell or recharge the battery
No confidence beep Out of order Return the equipment for maintenance
Displays. Steady audible and
visual alarms no resettable
No SMART BLOCK or bad contact
Defective SMART BLOCK
Turn off the instrument, then replace the
SMART BLOCK
Change the SMART BLOCK
Displays permanently or
alternately with the measurement.
Steady or unsteady audible and
visual alarm
Primary cell or rechargeable battery life
is ending
Change the primary cell or
recharge the battery
Displays
Bad calibration
SMART BLOCK life exhausted
Turn off the instrument
Recalibrate the SMART BLOCK
Change the SMART BLOCK
Displays. Steady audible and
visual alarm
Over measurement range Leave the area
Displays
Auto zero adjustment not performed Reinitiate auto zero or
change SMART BLOCK
Displays
Impossible to auto zero Stay in a safe atmosphere
Reinitiate the auto zero
Check the temperature 0 - 45°C
or
At turn on
Loss of time setting
Loss of memory
Change the lithium cell
Change the lithium cell
Measurements are negative Zero not adjusted Perform auto zero
Impossible to printout Bad connection Check installation
FAULT
BATTERY
USED CELL
OVERANGE
NO CAL
NOT POSS
C FAULT
C FAULT