ENMET Corporation TX2000 / OX2000
12
4.3.2 OX2000 Calibration
In OX2000 version, this adjustment can be done without gas but must be in clean air.
Enter the maintenance menu.
Press both switches at the same time for 5 seconds:
&
(During this time, measurement is displaying)
5 seconds later, the maintenance pictogram appears
Release the switches
At this time, the operator has 5 seconds to enter the
maintenance menus or the instrument returns to operation
mode.
· Then proceed with following
Press BACKLIGHT switch to toggle
between (PrG) and (CAL)
to continue See Sections:
4.2 for Programming
4.3.3 for Calibration
Figure 4: Entrance to Maintenance Menu Flow Chart
Do you want to reach
the menus?
N
O
YES
No action on switches
Return to normal operation,
5 seconds late
r
Press ALARM switch 3 times
Maintenance pictogram is flashing
You have access to:
Calibration (CAL) and
Programming (PrG) menus.