Spectrum ON-LINE ENMET Corporation
13
Table 2: Spectrum Calibration Voltage and Countdown Times
NOTE: The following are values used for production calibration. Not all gases are available for field calibration;
contact ENMET customer service personnel.
Gas Calibration Gas
* Calibration Cup ∆
Calibration Voltage Countdown Time
Ammonia
25ppm * 03620-011 0.67 180 sec
Arsine (2)
0.5ppm PH
3
= 0.38ppm
03620-018 ∆
0.10 120 sec
Carbon Monoxide
100ppm * 03620-011 1.70 120 sec
Chlorine (2)
5ppm
03620-011 ∆
1.40 180 sec
Fluorine (2)
3ppm Cl
2
= 7.5ppm
03620-011 ∆
1.80 240 sec
Hydrazine
0.3ppm
TBD
0.91 240 sec
Hydrogen
800ppm * 03620-011 2.20 180 sec
Hydrogen Chloride (2)
10ppm
03620-018 ∆
2.10 180 sec
Hydrogen Cyanide
10ppm * 03620-011 2.10 120 sec
Hydrogen Fluoride (1)
5ppm
TBD
1.40 240 sec
Hydrogen Sulfide
20ppm * 03620-011 0.46 120 sec
Nitric Oxide
25ppm
TBD
0.67 90 sec
Nitrogen Dioxide
5ppm * 03620-011 1.30 120 sec
Oxygen
20.9%(air) * 03620-011 2.80 120 sec
Ozone (1)
0.3ppm
03620-018 ∆
0.75 240 sec
Phosgene (1)
0.66ppm
03620-018 ∆
0.18 240 sec
Phosphine (2)
0.5ppm * 03620-011 0.13 120 sec
Silane
5ppm
TBD
0.54 120 sec
Sulfur Dioxide
10ppm
TBD
2.10 90 sec
NOTE: All specifications stated in this manual may change without notice.
TBD – To Be Determined
(1) Require use of Sample Draw System. See Figure 5.
(2) Recommended use of Sample Draw System, for enhanced sensitivity and speed of response. See Figure 5.
* Initial Calibration Cup 02552-008
For ALL Sensor Replacement
∆ Initial Calibration Cup 02552-023
For Sensor Replacement
Sample Adapter 03620-018
Calibration Cup 03620-011
Example of Sample Draw System
Figure 5: Identification of Calibration Adapters & Sample Draw System
To Sensor
Sample Draw System 04565-314
Shown with Spectrum OnLine attached
To Sensor
To Sensor
73073-020
quick release
Gas Cylinder
& Regulator
To Sensor